Model Number CI-1400-02H |
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
Therapeutic Effects, Unexpected (2099)
|
Event Date 10/25/2018 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing sound quality issues and open electrodes.Programming adjustment were made, however, the issue did not resolve.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
The external visual inspection revealed sliced silastic on the fantail region and at the straight portion of the array.In addition, the silastic overmold was damaged on both top and bottom covers.These are believed to have occurred during revision surgery.The photographic imaging inspection revealed several broken electrode wires at the fantail region.In addition, cut electrode wires observed on the fantail and the straight portion of the array.These are believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.The photographic imaging inspection and the scanning electron microscopy analysis revealed evidence of fatigue breaks on six electrode wires at the fantail region.These breaks can be associated with the sound quality and the amount of open electrodes reported.A corrective action was implemented.This is the final report.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.
|
|
Search Alerts/Recalls
|