• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Catalog Number 07028091190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
The customer questioned thyroid results for 3 patient samples tested for elecsys tsh, elecsys ft4 iii, and elecsys ft3 iii on a cobas e801 module.The 3 patient samples were submitted for investigation where discrepant results were identified for tsh and ft3 iii between the customer's e801 module, the siemens centaur method, an e801 module used at the investigation site, a cobas 8000 e 602 module used at the investigation site and a cobas e 411 immunoassay analyzer used at the investigation site.It is not known if any incorrect results were reported outside of the laboratory.This medwatch will cover tsh.Refer to medwatch with patient identifier (b)(6) for information on the ft3 iii results.There was no allegation of an adverse event.The serial number for the customer's e801 module was not provided.The e602 module serial number was (b)(4).The e411 analyzer serial number was (b)(4).The e801 module serial number used at the investigation site was (b)(4).The tsh reagent lot number used with the e602 module was 341699 with an expiration date of 28-feb-2019.The tsh reagent lot number used with the e411 analyzer was 328081 with an expiration date of 31-dec-2018.The tsh reagent lot number used with the e801 module at the investigation site was 331814 with an expiration date of 31-aug-2019.Calibration and qc at the investigation site were acceptable.All tsh results, with all types of analyzers, were near the lowest level of the normal reference ranges of the respective assays.Assays from different manufacturers, in this case siemens, can generate different results.From the information provided, a general reagent issue can be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Common Device Name
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7923043
Report Number1823260-2018-03375
Device Sequence Number1807596
Product Code JLW
UDI-Device Identifier04015630939831
UDI-Public04015630939831
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K961491
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07028091190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2018
Initial Report FDA Received Date10/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-