• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. DYNANITE NITI STAPLE W/INSTRS 18WX18L; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DYNANITE NITI STAPLE W/INSTRS 18WX18L; STAPLE, FIXATION, BONE Back to Search Results
Model Number DYNANITE NITI STAPLE W/INSTRS 18WX18L
Health Effect - Clinical Code Injury (2348)
Date of Event 03/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that the leg on the ar-8719ds-1818, dynanite nitinol staple implant system 18w x 18l is broken.The rep stated that it is unknown how or when the implant leg broke.The patient came back for a follow up visit and the device was discovered broken on the x-ray that was taken.The rep stated that the ar-8719ds-1818 remains implanted in the patient, and that there is not any record of patient discomfort.The rep reported that the surgeon is discussing a hardware removal surgery.Additional information requested regarding date of removal.Additional information received on 09/12/2018: the rep confirmed that the original procedure took place on (b)(6) 2018.The revision surgery has been scheduled for (b)(6) /2018.Additional information received on 09/14/2018: the rep confirmed that a hardware removal and revision evans procedure with illiac crest plate fixation took place on (b)(6) 2018.The revision was completed by the same surgeon who performed the original surgery, and took at the same facility where the original surgery was performed.The following arthrex part and lot numbers were explanted during the revision surgery on (b)(6) 2018: 1.Ar-8719ds-1818 / lot: 800980008.2.Ar-8942w-1810 / lot: 4210.The following arthrex part and lot numbers were implanted during the revision surgery on (b)(6) 2018: 1.Ar-8942r-l / lot: 10123947.2.Ar-8935-20 / lot: 10001844.3.Ar-8935-22 / lot: 10001427.4.Ar-8935l-20 / lot: 1336554.5.Ar-8935l-22 / lot: 191236.Additional information received on 09/20/2018: there were two arthrex representatives present during the revision surgery.The broken staple was removed with no new implant used in the same site.The rep confirmed that the bone quality was good.During the revision surgery the surgeon performed another procedure in the calcaneus using an h plate.An allosync 1 cc graft was used.
 
Additional Manufacturer Narrative
Complaint confirmed.Dimensional and material analysis results confirmed the ar-8719ds-1818 met all as-received conditions.Visual evaluation revealed that approximately 10.92mm of one of the staple legs had broken off of the device prior to receipt for investigation.This event is most likely caused by user mechanical damage to device, such as hitting the device with another device, prying/leveraging, or excessive bending forces applied during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNANITE NITI STAPLE W/INSTRS 18WX18L
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key7923115
Report Number1220246-2018-00684
Device Sequence Number1639860
Product Code JDR
UDI-Device Identifier00888867288140
UDI-Public00888867288140
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K172052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2019
Device Model NumberDYNANITE NITI STAPLE W/INSTRS 18WX18L
Device Catalogue NumberAR-8719DS-1818
Device Lot Number800980008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer09/10/2018
Initial Report FDA Received Date10/01/2018
Supplement Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-