| Catalog Number |
13200330 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
During gamma3 surgery, u-blade lag screw connector was broken in two at black marker portion when inserting a u-blade using u-blade inserter.The broken connector was completely removed from the patient.
|
| |
|
Event or Problem Description
|
|
During gamma3 surgery, u-blade lag screw connector was broken in two at black marker portion when inserting a u-blade using u-blade inserter.The broken connector was completely removed from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that u-blade lag screw connector gamma3® was alleged of instruments - broken, deformed, worn or scratched could be confirmed.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by improper handling.The device inspection revealed the following: appearance of the breakage surfaces in combination with slightly oblique fracture line indicates that the device broke in a forced brittle manner due to disproportional force and/or bending stresses by levering the instrument.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.
|
| |
|