| Catalog Number |
109.669 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Tissue Damage (2104)
|
| Date of Event |
07/10/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 1 month after the dental implant and bridge were placed in ada site 10 in the mouth, the implant did not integrate in this over prepped site and was removed.A different implant was then placed.The product will be forwarded to the manufacturer for investigation.There were no reported patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.The dentist reported that immediate load procedure was performed.
|
| |
|
Event or Problem Description
|
|
The clinician reported that 1 month after the dental implant and bridge were placed in ada site 10 in the mouth, the implant did not integrate in this overprepped site and was removed.A different implant was then placed.The product will be forwarded to the manufacturer for investigation.There were no reported patient complications.(b)(4).
|
| |
|