BAXTER HEALTHCARE - AIBONITO CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2N8378
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the clamp of the clear link catheter extension set sliced into the plastic tubing resulting in a hole.This occurred when the patient was sent to ct and the patient's iv was clamped.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph identified a hole in the tubing.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - AIBONITO
aibonito
MDR Report Key7924153
Report Number1416980-2018-06259
Device Sequence Number19228882
Product Code FPA
UDI-Device Identifier00085412048918
UDI-Public(01)00085412048918
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K112893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2023
Device Catalogue Number2N8378
Device Lot NumberUR18B23081
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/07/2018
Supplement Date Received by Manufacturer10/23/2018
Initial Report FDA Received Date10/01/2018
Supplement Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-