Catalog Number 71420162
Health Effect - Clinical Codes Pain (1994); No Information (3190)
Date of Event 05/14/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that revision surgery was performed.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and based on the available information, the root causes for the ¿lateral instability of the left patella, or the subsequent full revision surgery cannot be concluded.There has been no cause or reason revealed or speculated in any of the documents provided except the osteophytes and the possibility of osgood-schlatter¿s disease.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Event or Problem Description
It was reported that a revision surgery was performed.Several treatments were done such as analgesics and scan tests, which ended up in a revision surgery on (b)(6) 2018.Femoral, insert and tibial were removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEN II CEMENTED TIBIAL B/PLATE SZ 2 LEFT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH AND NEPHEW, INC.
1450 brooks road
memphis, tennessee
memphis 38116
MDR Report Key7924190
Report Number1020279-2018-02051
Device Sequence Number1913924
Product Code JWH
UDI-Device Identifier03596010206244
UDI-Public03596010206244
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71420162
Device Lot Number12KM16721
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/05/2018
Supplement Date Received by Manufacturer09/14/2018
09/14/2018
Initial Report FDA Received Date10/01/2018
Supplement Report FDA Received Date10/03/2018
12/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-