| Catalog Number |
71352701 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Radiation Exposure, Unintended (3164)
|
| Date of Event |
07/25/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported during a follow-up call, that patient had a drainage on (b)(6) 2016 due to bursitis.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices were not returned for evaluation.A clinical evaluation noted that all documents and images provided have been reviewed.However, the patient's dislocation cannot be confirmed.The two x-ray photos provided were from 2015 and they do not show the reported dislocation.Anaerobic/aerobic wound cultures were taken and showed no growth after 14 days.No further intervention has been planned.No further clinical/medical assessment is warranted at this time.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Should additional information be received, the complaint will be reopened.
|
| |
|
Additional Manufacturer Narrative
|
|
Two of the associated complaint devices were returned and evaluated.Visual examination of the oxinium head revealed several scratches and gouges to the lower flat surface of the head.One gouge was noted on the articular surface.The gouges and scratches most likely occurred during implant removal.The porous surface of the stem revealed visual evidence of on-growth.Further visuals of the stem surfaces exhibited scoring/gouges of the distal stem as well scratches to the femoral neck area.Also noted were areas of the porous region that had been scraped off.The depth and number of scratches suggest some amount of force was required to remove this implant.Visually, there is nothing to conclude the implant was the root cause of the issue for which this implant was revised.No material or manufacturing deviations were observed.The clinical/medical team concluded that although it has been communicated that a revision surgery has been performed, the revision op notes have not been provided.However, the newly provided documents have been reviewed and do not contribute to the medical investigation.Therefore, no further assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the complaint history revealed no prior complaints for the listed batch.Without the actual product involved our investigation cannot proceed.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.If the devices are received in the future, this complaint can be re-opened.
|
| |
|
Search Alerts/Recalls
|
|
|