• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORTRAK* 2 ENTERAL ACCESS SYSTEM; DH CORTRAK (EAS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORTRAK* 2 ENTERAL ACCESS SYSTEM; DH CORTRAK (EAS) Back to Search Results
Model Number 20-0950
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Death (1802); Low Oxygen Saturation (2477)
Type of Reportable Event Death
Additional Manufacturer Narrative
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 01-oct-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).The device was not returned.
 
Event or Problem Description
A medwatch (mw) report, fda report number fda 3500 form (b)(4), was received.It was reported that a feeding tube was placed using the cortrak 2 system.Wave form the same as wave placement from 2 days earlier.Kidney, ureter, and bladder (kub) x-ray obtained due to cortrak being only at 45 mark to make sure it was in the stomach.A doctor told bedside nurse it was ok to use.When bedside nurse went to use tube, there was a large amount of blood coming from the tube.Had secretary call radiology to get a stat reading on the x-ray.The tube was removed in the meantime radiology md stated the tube was in the right lung.The pt began to decline with decreasing blood oxygen saturation (spo2) and eventually to the point a code blue was called, during the very lengthy code, the husband was called and he said to stop the code the code blue was stopped and pt expired.Another doctor was present for the code.When asked he said not to call the medical examiner as it was not needed.
 
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 25-oct-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 1605001, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.One (1) cortrak eas unit was returned.The monitor unit was inspected for damages and breakages, there were no breakages or damages found on the blue housing and orange button.The device was evaluated by connecting the returned monitor unit, smart receiver unit (sru) interconnect cable to the lab's anatomical model and transmitting stylet.Functional testing of the unit indicated that the unit works as intended.After the setup, verification of the cortrak system and the completion of the initialization, the functionality of the unit performed as intended.The complaint could not be confirmed because the placement tracings were not present.The root cause was not identified.All information reasonably known as of 11-dec-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
DH CORTRAK (EAS)
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key7924738
Report Number3006646024-2018-00034
Device Sequence Number1787217
Product Code KNT
UDI-Device Identifier10815149024582
UDI-Public10815149024582
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K113351
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number20-0950
Device Catalogue Number104720100
Device Lot Number1605001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer10/03/2018
11/13/2018
Initial Report FDA Received Date10/01/2018
Supplement Report FDA Received Date10/26/2018
12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age70 YR
-
-