• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751PNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751PNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751PNAL
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had keypad anomaly and failed battery.The customer¿s blood glucose level was unknown.The customer reported that they had diminished battery life, rejected new battery a few times, buttons sticking and require multiple presses causing wrong.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device received with all buttons responding properly.No damage on keypad assembly.No unlocked lcd keypad connector.Device received with all operating currents within specification and passed a21 error test and displacement test.No unexpected failed battery alarm anomalies noted.No unexpected low battery alarm anomalies noted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751PNAL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7924990
Report Number2032227-2018-35808
Device Sequence Number16711780
Product Code OZO
UDI-Device Identifier00643169507333
UDI-Public(01)00643169507333
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-751PNAL
Device Catalogue NumberMMT-751PNAL
Device Lot NumberA3751PNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2019
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer11/22/2019
Initial Report FDA Received Date10/02/2018
Supplement Report FDA Received Date11/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient Weight195
-
-