| Model Number |
1011-0139-05 |
| Medical Device Problem Code |
Energy Output Problem (1431)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during training by a zoll medical representative, the associated defibrillator's defib output was out of specification using these internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The internal paddles were returned to zoll medical corporation and the customer's report was not replicated or confirmed.The internal paddles were put through extensive testing including multiple 2 joule and 30 joule tests without duplicating the reported malfunction.The internal paddles were inspected with no discrepancies found.The internal paddles were scrapped and the customer received replacement paddles.
|
| |
|