| Model Number |
FG-06551-001A |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The root cause of the incident is unknown at the time of the initial report.The complaint device was not returned for evaluation yet, therefore the investigation and root cause analysis could not be performed.The dhr review of the complaint lot 645007 and related subassemblies lots 646116 and 646115 was performed, there was no deviation found in the records.The devices passed all 100% in process and aql based qc inspections.The manufacturer informed manufacturing operators and qc inspectors about the complaint.No remedial actions were initiated at the time of initial report due to the suspected user error (withdrawal of introducer when resistance was felt) and / or the incident could be related to the patient condition/anatomy based on results of investigations performed within the previous complaints.Additionally, a vessel spasm (which could be a potential cause of the resistance) can occur during the procedure, as defined under the potential complications section in ifu.(b)(4).
|
| |
|
Event or Problem Description
|
|
The manufacturer was informed, that during removal of the sheath, there was a resistance.The sheath started to elongate and then came apart about 10 cm from the hub.The artery forceps were used to remove the rest of the sheath and nothing was left in the patient.
|
| |
|