|
Pc-(b)(4).It was reported performance breakage suture.A needle-suture piece and two unopened samples were returned for analysis.During the visual inspection of opened sample (needle/suture pieces, the swage and attachment area were as expected, and the ends suture piece was observed busted.Also, a functional test was performed and the tensile force were above the minimum requirements.Two unopened samples were inspected visually, no defects were found on the packages.The samples were opened, and the swage and attachment area of needles were as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were observed.A functional test was performed and the tensile force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Due to the condition of opened sample, the assignable cause of performance breakage suture, was suggests an improper handling of the sample.Per the condition of the representative samples, no performance - breakage suture was found on the sutures and the tested samples met the finished goods requirements.Representative samples returned to support investigation of 5 device issues captured in reports 2210968-2018-76158, 2210968-2018-76159, 2210968-2018-76160, 2210968-2018-76161, 2210968-2018-76162.Analysis results have been included in follow-up reports for each.
|