| Catalog Number |
112240 |
| Medical Device Problem Codes |
Break (1069); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Adapter tightening bolt broke off in 3 pin clamp.Case type: tha.
|
| |
|
Event or Problem Description
|
|
Adapter tightening bolt broke off in 3 pin clamp.Case type: tha.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: it was reported that adapter tightening bolt broke off in 3 pin clamp.Product evaluation and results: the product was unavailable for inspection as the product was not returned.Product history review: review of the product history records indicate 46 devices were manufactured under lot no 19130711 and accepted into final stock on (b)(6) 2011.No non-conformances were identified during inspection.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 112240, lot number 19130711, shows 01 additional complaints related to the failure in this investigation.Complaint (b)(4).Conclusions: the failure mode could not be confirmed because the part was not available for evaluation.If device and/or additional information become available, this investigation will be reopened.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
|
| |
|