• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH CENTRAX DURATION 26MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH CENTRAX DURATION 26MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6226-2-650
Medical Device Problem Code Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Unable to lock the ring while removing the green locking ring and using the centranx forceps to attach it to the head.As a result of confirmation, a part of polyethylene was broken.Used another size cup.
 
Additional Manufacturer Narrative
An event regarding crack/fracture of a centrax head during implantation was reported.The event was not confirmed.Conclusion: three photographs of the component were provided for review.Damage consistent with attempted assembly is observed on the poly of the head but the reported fracture of the poly is not visible, therefore not confirmed.The exact cause of the event could not be determined because insufficient information was provided.Additional information, including evaluation of the device and the operative report are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
Unable to lock the ring while removing the green locking ring and using the centranx forceps to attach it to the head.As a result of confirmation, a part of polyethylene was broken.Used another size cup.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRAX DURATION 26MM X 50MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7926670
Report Number0002249697-2018-03143
Device Sequence Number1346775
Product Code JDI
UDI-Device Identifier07613327032239
UDI-Public07613327032239
Combination Product (Y/N)N
PMA/510(K) Number
K972792
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number6226-2-650
Device Lot Number1T6KW1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer11/14/2018
Initial Report FDA Received Date10/02/2018
Supplement Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age38 YR
Patient Weight70
-
-