No medical records and no medical images were provided to the manufacturer.The lot number was not provided, a review of the device history records has not been performed.The device has been returned for evaluation.The investigation is currently underway.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
A complete manufacturing review could not be conducted, as the lot number was not reported for this device.The result of the investigation is confirmed for the failure mode reported, device incompatibility.During evaluation the guidewire could not pass through the catheter due to the distal markerband been squashed.A longitudinal balloon rupture and tip break were also noted.The source of these defects cannot be proven but it possibly due to user techniques during the procedure.It is unlikely that the device incompatibility issue is manufacturing related as a 100% guidewire check is performed during catheter production.A complete manufacturing review could not be conducted, as the lot number was not reported for this device.The definitive root cause for the device incompatibility issue could not be determined based upon available information.It is unknown whether patient factors, handling or procedural techniques may have contributed to the reported event.Ifu for sleek pta rapid exchange (rx) dilatation catheter product family was reviewed the following sections are applicable: the sleek percutaneous transluminal angioplasty (pta) peripheral catheter family comprise a range of sizes of rapid exchange catheters for peripheral angioplasty.The catheter¿s proximal tubing is 304v stainless steel and the distal coaxial tubings are nylon co-polymer blend.The lumen of the shaft is used for the purpose of inflating and deflating the balloon.A second lumen at the tip is used for advancing the guidewire.The sleek catheters are intended for balloon dilatation of the femoral, popliteal and infra-popliteal arteries.These catheters are not designed to be used in the coronary arteries.Reuse, resterilization, reprocessing and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Carefully inspect the catheter prior to use to verify that the catheter has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.Proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.If the hypotube kinks prior to or during use the catheter should be discarded.No attempt should be made to straighten a kink in the hypotube.Store in a cool, dark, dry place.Use the catheter prior to the ¿use by¿ date specified on the package.Directions for use: inspection and preparation remove the balloon sleeve by first withdrawing the shipping mandrel slightly and then slowly removing the sleeve while holding the catheter as close to the balloon as possible.If any resistance is felt, or if any stretching of the catheter is observed while removing the balloon sleeve, the product should not be used.The catheter should then be inspected for bends, kinks or stretched portions.Do not use if product damage is evident.Simultaneously withdraw the dilatation catheter and guidewire from the guiding catheter/sheath.As the balloon exits the vessel, use a smooth, gentle, steady, motion.If resistance is felt upon removal then the balloon, guidewire and the guiding catheter/sheath should be removed together as a unit under fluoroscopic guidance, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and guiding catheter/sheath as a unit and withdrawing both together, using a gentle twisting motion combined with traction.(b)(4).
|