MAKO SURGICAL CORP. FEMORAL TRIAL STD LM/RL SIZE 4; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
|
Back to Search Results |
|
| Catalog Number |
170504 |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An event regarding damage involving a mako trial was reported.The event was confirmed following visual inspection.Method & results: device evaluation and results: visual inspection: a review by a materials engineer noted: "damage observed on the threads and anterior distal surface of the condyle consistent with contact against hard object".Material analysis: a review by a materials engineer noted: "examination of the returned device with engineer indicated damage observed on the threads and anterior distal surface of the condyle consistent with contact against hard object.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined".Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: examination of the returned device with engineer indicated damage observed on the threads and anterior distal surface of the condyle consistent with contact against hard object.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|
Event or Problem Description
|
|
Surgeon spliced the teeth of the lm size 4 femoral trial by impacting the slap hammer.
|
| |
|
Search Alerts/Recalls
|
|
|