Catalog Number ATG80186
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 09/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure of an in stent re-stenosis, the pta balloon allegedly detached from the catheter shaft as the health care provider was cleaning the balloon before inserting into the patient for the second inflation attempt.Another balloon was prepped and used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found the a complete break at the proximal balloon weld.The distal weld was intact.Evidence of material transfer was noted at the weld location between the balloon and the shaft.Therefore, the investigation is confirmed for a weld break and is unconfirmed for the reported balloon detachment.The definitive root cause for the identified weld break could not be determined based upon the available information.It is unknown if procedural issues or the original sheath used in the procedure contributed to the reported event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during an angioplasty procedure of an in stent re-stenosis, the pta balloon allegedly detached from the catheter shaft as the health care provider was cleaning the balloon before inserting into the patient for the second inflation attempt.Another balloon was prepped and used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7927846
Report Number2020394-2018-01825
Device Sequence Number1346693
Product Code LIT
UDI-Device Identifier00801741060892
UDI-Public(01)00801741060892
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K122984
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2021
Device Catalogue NumberATG80186
Device Lot Number93TC0139
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Initial Date Received by Manufacturer 09/06/2018
Supplement Date Received by Manufacturer01/22/2019
Initial Report FDA Received Date10/02/2018
Supplement Report FDA Received Date01/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-