WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOOK/SCREW HOLDER WITH 4.0MM HEX; MISC ORTHO SURGICAL INSTRUMENT Back to Search Results
Model Number 388.612
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information is not available for reporting.Date of event is unknown.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.A review of the device history records has been requested and is currently pending completion.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, during spinal fusion surgery, the tip of (4) universal spinal system (uss) hook and screw holders got bent during screw manipulation.They screw on to the screw heads and the surgeon manipulated them for rod and cap insertion.As they were being handled, the surgeon noticed that some were bending.It is unknown if there was any surgical delay.Patient outcome was unknown.This complaint involves one (1) hook/screw holder with 4.0mmhex devices.This report is 4 of 4 for (b)(4).
 
Event or Problem Description
The procedure occurred on (b)(6) 2018.Concomitant devices: screw (part: unknown, lot: unknown, quantity: unknown), rod (part: unknown, lot: unknown, quantity: unknown), cap (part: unknown, lot: unknown, quantity: unknown).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Complainant part is not expected to be returned for manufacturer review/investigation.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOOK/SCREW HOLDER WITH 4.0MM HEX
Common Device Name
MISC ORTHO SURGICAL INSTRUMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7928337
Report Number2939274-2018-54103
Device Sequence Number19316956
Product Code LXH
UDI-Device Identifier10705034773629
UDI-Public(01)10705034773629
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number388.612
Device Catalogue Number388.612
Device Lot Number4523211
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/07/2018
Supplement Date Received by Manufacturer10/26/2018
11/22/2018
Initial Report FDA Received Date10/02/2018
Supplement Report FDA Received Date10/26/2018
11/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-