| Model Number |
MPX5302-C |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer reported a leak occurred during patient use.No patient harm was reported.
|
| |
|
Event or Problem Description
|
|
The customer reported a leak occurred during patient use.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of a leak was confirmed.Functional testing was performed; a leak was observed at the attachment of the maxplus valve to the female luer.Magnified inspection showed that the source of the leak was a thin crack spanning the length of the female luer body.The root cause of the damage was not determined.
|
| |
|