SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 36 MM M/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 71343604
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 09/05/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the cpcs r3 revised for dislocation.Hip stable on table.Impinge anterior acetabular osteophyte.Head stem liner & screws removed.Polar cup cemented into shell.Smaller stem cemented.Polar articulation stable.Nil implants for retrieval.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned for evaluation.Our investigation including a review of complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A clinical analysis noted that all documents and/or images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical investigation.Based on insufficient information, a thorough clinical assessment cannot be performed at this time.Without the return of the actual products involved, our investigation of this report is inconclusive.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 36 MM M/+4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7928812
Report Number1020279-2018-01969
Device Sequence Number16622802
Product Code JDI
UDI-Device Identifier03596010463296
UDI-Public03596010463296
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K022958
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71343604
Device Lot Number17KM02029
Initial Date Received by Manufacturer 09/05/2018
Supplement Date Received by Manufacturer09/05/2018
Initial Report FDA Received Date10/02/2018
Supplement Report FDA Received Date11/29/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-