MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAS
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had keypad anomaly.The customer¿s blood glucose level was unknown.The customer reported that the button wasn't working on the pump.The customer reported that they had damage on the screen and was not able to read the pump.The customer was advised to discontinue use of the pump and revert to backup plan per health care professional instructions.The customer was advised the pump will need to be replaced.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed self test.Insulin pump received with minor scratched lcd window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7928984
Report Number2032227-2018-36304
Device Sequence Number11262193
Product Code OYC
UDI-Device Identifier00613994622655
UDI-Public(01)00613994622655
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAS
Device Catalogue NumberMMT-522NAS
Device Lot NumberA6522NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2018
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer01/08/2019
Initial Report FDA Received Date10/02/2018
Supplement Report FDA Received Date01/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight135
-
-