Catalog Number 314.118
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no patient involvement associated with the complained event.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the t25 stardrive screwdriver was noticed to be loose and the shaft portion disconnected or broke off from the handle.The issue was identified at the back table when the surgical technician was preparing for the case while the operating room (or) team was positioning the patient.The procedure was a simple implant removal of a couple screws.The procedure was completed uneventfully without any delay.The issue did not impact the procedure or the patient.There was no patient involvement reported.This report is for one (1) stardrive screwdriver t25.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device evaluated by mfr: part: 314.118; synthes lot: 4905140; supplier lot: na; release to warehouse date: december 10, 2004; manufactured by synthes brandywine.The raw material was confirmed to be correct per the specification with no (relevant) non-conformance noted.No nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Flow: broken visual inspection: the stardrive screwdriver was received at us cq with minimal signs of wear.The screwdriver shaft and the handle were still together but easily pulled apart.The screwdriver shaft connection to the handle was loose.This agrees with the reported complaint condition; thus, the complaint is confirmed.Dimensional inspection: dimensional analysis was completed, the outer diameter of the screwdriver shaft measured 6.99 mm.This is within specification of 6.97 mm to 7 mm, based on drawing.Document/specification review: the following drawing(s) was reviewed; t25 stardrive¿ screwdriver self-retaining 290 mm; screwdriver shaft self-retaining/ t25 stardrive¿.The raw material was confirmed to be correct per the specification with no (relevant) non-conformance noted.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Conclusion: the complaint condition is confirmed as the stardrive screwdriver was received with the shaft and the handle connection loose and easy to come apart.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces (such as being dropped, struck off axis or damaged during sterile processing).Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARDRIVE SCREWDRIVER T25
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7929703
Report Number2939274-2018-54111
Device Sequence Number1797571
Product Code HXX
UDI-Device Identifier10886982188557
UDI-Public(01)10886982188557
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.118
Device Lot Number4905140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer10/11/2018
10/11/2018
Initial Report FDA Received Date10/03/2018
Supplement Report FDA Received Date10/26/2018
10/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-