MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC COMPRESSION FORCEPS; INSTRUMENT COMPRESSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC COMPRESSION FORCEPS; INSTRUMENT COMPRESSION Back to Search Results
Model Number 03.211.400
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device history records review was completed for part# 03.211.400, lot# t950802.Manufacturing location: (b)(4), release to warehouse date: jan 28, 2011.A review of the device history records showed that there were no issues at the time of manufacturing of this device and it's sub components that would contribute to the complaint condition.The raw material certificate was reviewed and the used material was according to the specification of the device.No non conformance reports were generated during production.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Product investigation was completed.Upon inspection it was noticed that a piece of the small flap had chipped off and contributed to the complaint condition.Functional test was able to replicate the complaint condition.It is likely that over seven years of routine use contributed to the piece of the small flap chipping off.Due to post-market damage, and the damaged area being to small to access, an effective dimensional inspection could not be performed.Relevant drawings were reviewed during investigation, and determined to be suitable for the intended design, application, and dimensional conformity when used and handled as recommended.Although a definitive root cause could not be determined, it is most likely that the complaint condition is due to routine use/processing over its over seven-year lifespan, and/or rough usage.There were no issues during the manufacture of this product that would contribute to this complaint condition.During the investigation, no product design issues or discrepancies were observed that may have contributed to the complaint condition.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that a returned compression forceps from fsl facility was noted as ¿not locking¿.The device was inspected as part of the routine incoming inspection on (b)(6) 2018.The forceps was being out of specification.There was no patient or hospital involvement.Upon inspection it was noticed that a piece of the small flap had chipped off and contributed to the complaint condition.This report is for one (1) compression forceps.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPRESSION FORCEPS
Common Device Name
INSTRUMENT COMPRESSION
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7929755
Report Number2939274-2018-54113
Device Sequence Number19169013
Product Code HWN
UDI-Device Identifier10886982076267
UDI-Public(01)10886982076267
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.211.400
Device Catalogue Number03.211.400
Device Lot NumberT950802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/02/2018
Initial Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-