PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control provided the customer with technical assistance and provided the part number for the main keypad assembly.The device was not returned to physio-control for an evaluation.The cause of the reported issue could not be conclusively determined.Device not evaluated by manufacturer.
 
Event or Problem Description
The customer, a commercial distributor, contacted physio-control to report that their device had logged an event code in its memory code which is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
The customer advised physio-control that the main keypad was replaced in the device and this resolved the reported issue.The removed keypad was not returned to physio for evaluation.Based on the information available, physio-control has determined that it¿s reasonable to conclude that the likely cause of the reported issue was that the shock switch located on the main keypad assembly was out of tolerance due to excessive resistance.When this issue occurs, an event code will be logged in the device's memory following a failure of the device to deliver defibrillation energy.
 
Event or Problem Description
The customer, a commercial distributor, contacted physio-control to report that their device had logged an event code in its memory code which is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
This product complaint requires a supplemental mdr to document that field action number 301587611/18/2019-001c is relevant to the reported issue.
 
Event or Problem Description
The customer, a commercial distributor, contacted physio-control to report that their device had logged an event code in its memory code which is related to a potential failure of the device to deliver defibrillation energy.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7930123
Report Number0003015876-2018-01630
Device Sequence Number1798159
Product Code MKJ
UDI-Device Identifier00883873911761
UDI-Public00883873911761
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001243
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer11/27/2018
11/18/2019
Initial Report FDA Received Date10/03/2018
Supplement Report FDA Received Date11/28/2018
11/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3015876-11/18/2019-001C
Patient Sequence Number1
-
-