• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 9588
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is a combination product.
 
Event or Problem Description
It was reported that stent dislodgment occurred.The target lesion was located in the mildly tortuous and mildly calcified left circumflex (lcx) artery.After a bsc stent was deployed in the proximal lcx, a 2.25x32mm promus premier stent was advanced but failed to cross the distal lesion.During removal, the stent got caught on the distal edge of the proximal stent and dislodged.Subsequently, the dislodged stent was deployed using a semi compliant and non compliant balloons and the procedure was completed.No patient complications were reported and the patient's status was fine.The patient was discharged the following day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key7930666
Report Number2134265-2018-61555
Device Sequence Number13815356
Product Code NIQ
UDI-Device Identifier08714729844945
UDI-Public08714729844945
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110010
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/08/2019
Device Model Number9588
Device Catalogue Number9588
Device Lot Number0021286295
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2018
Initial Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-