SMITH AND NEPHEW, INC. JRNY BCS II FEM COCR LT SZ 5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 74024925
Health Effect - Clinical Code Pain (1994)
Date of Event 12/18/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed on the patient due to pain in the knee.No evidence of infection was found.Laxity of the components was present most notably in full flexion.Femoral, tibial and articular devices were explanted.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.The clinical/medical team concluded, this case reports a revision surgery was performed on the patient due to pain in the knee.The date of the initial implantation is unknown.It was further communicated that no evidence of infection was found and laxity of the components was present most notably in full flexion.Details regarding the patient¿s weight-bearing status, medical history, bone quality, and other additional clinical relevant information have not been provided.Based on the limited information provided and without the return of the device for evaluation and diagnostic images, the root cause of the reported pain and laxity of the joint cannot be determined.Should additional information becomes available this issue can be re-elevated.No further clinical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY BCS II FEM COCR LT SZ 5
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH AND NEPHEW, INC.
1450 brooks road
memphis, tennessee
memphis 38116
MDR Report Key7930753
Report Number1020279-2018-01984
Device Sequence Number2064169
Product Code JWH
UDI-Device Identifier00885556611432
UDI-Public00885556611432
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/24/2025
Device Catalogue Number74024925
Device Lot Number01511883
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/05/2018
Supplement Date Received by Manufacturer09/05/2018
Initial Report FDA Received Date10/03/2018
Supplement Report FDA Received Date06/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age48 YR
-
-