| Catalog Number |
0250070601 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event.There was no associated procedure and no reports of adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the customer reported that after installing the battery, the device was put on charge.After a few minutes smoke began to come out of the device.It was quickly disconnected.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be a possible power surge via power connection to the unit and/or an incorrect connection configuration between the battery and the unit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event.There was no associated procedure and no reports of adverse consequences.
|
| |
|