| Catalog Number |
3105000801 |
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
07/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device manufacture date is not known at this time.However, should it become available, it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that post op pain occured as a result of prolonged traction.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: patient experienced post op pain due to prolonged traction probable root cause: design: boot does not adequately distribute pressure.Boots are not adequately secured.Boas don¿t provide sufficient tactile feedback.No indication of force felt by patient.Process: boa laces routed incorrectly.Use errors.The product was not returned for investigation therefore the reported failure mode was not confirmed.The reported failure mode will be monitored for future reoccurrence.The device manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that post op pain occured as a result of prolonged traction.
|
| |
|
Event or Problem Description
|
|
It was reported that post op pain occured as a result of prolonged traction.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation therefore the reported failure mode was not confirmed.Alleged failure: patient experienced post op pain due to prolonged traction.Probable root cause: design: materials bend/dent under force.Design of joints and locking mechanisms.Connectors or locks break under traction force.Casters slip on floor.Boot spine bends.Boot boas unreel due to torque limit.Boa connection guide breaks under high tension.Patient position pad slides on table or patient slides on top of pad.Patient position belt not strong enough to prevent patient from sliding.Use error.The device manufacture date is not known.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Search Alerts/Recalls
|