Consumer reported complaint for high and erratic blood glucose test results.The customer is concerned with tests results of 263, 275, 195, 209 and 229 mg/dl.The customer's expected fasting blood glucose test result range is 100 - 150 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer non-fasting and produced test results of 242 mg/dl and 237 mg/dl using truemetrix air meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 10/23/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history (customer could not really see the date/time, and so he provided the information he thinks he sees; also customer stated that he has never set up the date/time): result 1 :263mg/dl date: (b)(6) 2018 time: 01:00pm fasting; result 2 :275mg/dl date: (b)(6) 2018 time: 01:00pm fasting; result 3 :195mg/dl date: (b)(6) 2018 time: 03:00pm fasting; result 4 :209mg/dl date: (b)(6) 2018 time: 04:00am fasting; result 5 :229mg/dl date:undisclosed time: undisclosed fasting.
|