| Catalog Number |
48230141 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of the investigation.Device remains implanted in patient.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient originally had an l3-s1 fusion and was brought back for an extension of metal work procedure back in (b)(6) 2018.The patient's l3 screws were removed and the surgeon put 2 more screws into l2 and connected the existing rods to 2 new rods with 2 side to side connectors.They were hand tightened as per surgical technique.The patient has come back for a follow up (completely asymptomatic) and an x-ray indicated that the connector has become loose and slid down the rod and rotated.No surgery is planned currently.There are no reports of adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: x-ray review, nc/capa history review, labelling review, risk assessment.Result: the customer reported event was confirmed via x rays.Rod has moved but hasn't disengaged from the connector.This was shown in the x rays as the right side connector has slid down the rod and appeared to have rotated.The device still remained implanted in the patient and lot # was not provided, therefore product history review could not be performed.Conclusion: as the device has not been returned a definite root cause cannot be determined.Possible root causes include surgeon over torqueing or under torqueing the blockers, damaged blockers on the connector, and/or excessive post op activity.Device still remained implanted in the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient originally had an l3-s1 fusion and was brought back for an extension of metal work procedure back in (b)(6) 2018.The patient's l3 screws were removed and the surgeon put 2 more screws into l2 and connected the existing rods to 2 new rods with 2 side to side connectors.They were hand tightened as per surgical technique.The patient has come back for a follow up (completely asymptomatic) and an x-ray indicated that the connector has become loose and slid down the rod and rotated.No surgery is planned currently.There are no reports of adverse consequences to the patient.
|
| |
|