| Model Number |
ZCT400 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If explanted; give date: n/a (not applicable).The lens remains implanted.The intraocular lens (iol) is not returning for evaluation as it remains implanted; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that model zct400 tecnis toric iol (intraocular lens) was initially placed at 84 degrees in her left eye (os).At 6-month study, iol rotation was measured at 94 degrees.This outcome does not significantly affect daily activities for the patient.No lens repositioning has been planned or performed and the lens remains implanted.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|