Additional Manufacturer Narrative
|
|
Date of event reported as 2018, exact date is not known.Additional product codes: hwc, jdq.A review of the device history records has been requested.Subject device has been received and is currently in the evaluation process.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure on an unknown date in 2018, a piece of a titanium sternal locking star plate broke off while being bent by two (2) universal bending pliers.It is unknown if there was a surgical delay.Procedure and patient outcomes are unknown.Concomitant devices: universal bending pliers (part 391.963.96, lot number unknown, quantity 2).This report is for one (1) sternal locking star plate.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record (dhr) review was conducted: device history record (dhr) review for top level assembly part#460.036 lot#4309446 manufacturing location: synthes monument released to warehouse date: 21.Aug.2001.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.12 hole slot component part# 460.036.1 lot# 4286494.The material was determined to be conforming with no relevant nonconformances at the time of manufacture based on review of the raw material dhr.12 hole tab component part# 460.036.2 lot# 4290900: the material was determined to be conforming with no relevant nonconformances at the time of manufacture based on review of the raw material dhr.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.No relevant ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.A product investigation was conducted.Visual inspection: upon inspection the complaint condition was confirmed as two of the outer plate holes of the tab half of the returned 12 hole plate.The received condition does agree with the complaint description.The reported complaint description indicated that the plate broke during contouring.Dimensional inspection per relevant drawing: outer plate hole thickness specification: 1.6mm ± 0.1.Document/specification review: relevant drawings were reviewed, and no product design issues or discrepancies were observed.Material review or hardness review: dhr review indicated that the returned plate was manufactured with a conforming raw material composition.Investigation conclusion: this complaint is confirmed.The reported complaint description indicated that the plate broke during contouring.It is possible that the user used excessive force or bent the plate excessively, which could have contributed to the complaint condition.During the investigation no unidentified product design/manufacturing issues or discrepancies were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|