| Model Number |
MMT-7008A |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The patient¿s blood glucose was 187 mg/dl and the sensor glucose was 40 mg/dl at the time of the incident.Difference between blood glucose and sensor glucose was 147 mg/dl.Insulin delivery was suspended due to sensor glucose values.Sensor glucose value that triggered the suspend event was unknown.Threshold/low suspend limit in sensor settings was 60 mg/dl.Customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The device is expected to be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened and used sensor and performed continuity resistance test and sensor passed per specifications.Performed the sensor initialization test using a new lab transmitter with a paradigm pump, connected the sensor to the transmitter and placed it in 100 buffered test solution, confirmed the communication icon was displayed and that the transmitter was flashing while connected to the sensor.Sensor passed per specifications.Cannot perform buffered test solution test as the sensor is beyond its expiration/event date.
|
| |
|
Search Alerts/Recalls
|