Product complaint # pc-(b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Investigation summary two empty open foils, two empty winding former two dispensed needle suture combinations, a needle suture piece and three unopened samples of product code (b)(4) , lot mdq792 were returned for analysis during the visual inspection of a needle/suture piece, the swage and attachment area were noted to be as expected and the end suture was cut that appears to be by use of surgical instrument.Also, body fluids were observed and the functional test (measured barbs) cannot be performed.During the visual inspection of five samples (two open and three unopened), the swage and attachment area were noted to be as expected, loops and the barbs were present on the strands.Ten barbs were measured in each sample and was noted a minimal depth, resulting in a smaller barb.No other defects were found.The barb to center of loop and first barb to end of needle barrel were inspected, and no barbs were noted, and this condition is correct according to ethicon requirements.According to condition of the needle-suture piece, the assignable cause of barb non engagement in tissue could not be determined.Due to the condition of five samples, the evaluation found that the barb depth was out of specification.The information provided is compiled monitored and reviewed on a routine basis for any associated trends.
|
(b)(4).Corrected device evaluated by mfr: five samples with lot # mdq792 were received for analysis.During the visual inspection, all barbs meet the specification requirement.The manufacturing records were reviewed and the manufacturing criteria were met prior to the release of this batch.According to the samples conditions, the samples met the finish goods requirements.Corrected information: adverse event problem.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Representative samples returned to support investigation of 3 device issues captured in reports 2210968-2018-76178, 2210968-2018-76173.Analysis results have been included in follow-up reports for each.
|