| Catalog Number |
352894 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
09/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device involved has been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: event 2: it was reported that upon administering versed and fentanyl, during a wisdom tooth extraction, blood was coming back up the iv tubing to the distal y-site, and blood leaked onto the floor.The procedure was actually completed when the blood leak was noticed.The patient did not suffer any adverse effects and did not require any additional intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Event 2: this report has been identified as b.Braun medical internal report number (b)(4).Twenty-seven (27) unopened samples were returned for evaluation.The returned sets were tested for leakage per specification.One (1) out of 27 sets was found to be leaking from the sonic weld of the distal safeday ultrasite valve.The remainder sets were found acceptable.The reported defect of leakage was confirmed through physical testing.The cause of the leakage was due to a weak sonic weld.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.All available information regarding this occurrence has been forwardedto our material review broad (mrb) business unit leaders and all personnel involved in the manufacturing and inspection of this product in order to heighten awareness.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|