• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US AML HIGH OFFSET LG STAT 13.5MM; AML IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US AML HIGH OFFSET LG STAT 13.5MM; AML IMPLANT : HIP FEMORAL STEM Back to Search Results
Catalog Number 155403135
Medical Device Problem Codes Corroded (1131); Degraded (1153); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Host-Tissue Reaction (1297); Pain (1994); Injury (2348); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); Not Applicable (3189)
Date of Event 03/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Reporter is an attorney.
 
Event or Problem Description
The patient was revised to address disassociation of the metal liner from the cup and related pain.Update ad 10 sep 2018: litigation alleges elevated blood metal levels and injury which inhibited the patient's ability to move.The law firm and lawyer were added to this complaint.An impacted product record for the stem and head was added due to the alleged elevated metal ion levels.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
In addition to what was previously alleged, ppf alleges pseudotumor.Added expiration date and dob.After review of medical records, patient was revised on (b)(6) 2017 due to mechanical complication after right hip replacement.It was noted that the taper junction had mild corrosion on the male portion of the taper, minimal corrision on the female taper.The head had moderate striped wear.Screw was prominent about 0.5mm.Liner was loose/dissociated.Added screw to the complaint.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  .
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AML HIGH OFFSET LG STAT 13.5MM
Common Device Name
AML IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7933611
Report Number1818910-2018-71269
Device Sequence Number19245884
Product Code LPH
UDI-Device Identifier10603295058441
UDI-Public10603295058441
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number155403135
Device Lot NumberC3DAX1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer03/05/2019
08/12/2019
12/31/2019
Initial Report FDA Received Date10/04/2018
Supplement Report FDA Received Date03/13/2019
09/06/2019
12/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-