| Catalog Number |
C8026 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Visual examination of the returned used device, item c8026, confirms the reported problem and found the nitinol wire misused, bent and broken off at the etched line.The broken piece was also returned for evaluation.This is a technique dependent device and the most likely cause of this issue is user related.The manufacturing documents from a device history record have been reviewed with special attention to the manufacturing and inspection of the product.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 2 complaints, regarding 2 devices, for this device family and failure mode.During the same time frame 26,676 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be.00007 per the instructions for use the user is advised the following; instrument is designed for use by surgeons experienced in the appropriate specialized procedures.It is the responsibility of the surgeon to become familiar with the proper techniques for use.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The customer reported that the 5129491, bioscrew hyperflex guidewire, broke during surgery.This caused a 2 hour delay, there was no reported patient injury, however, the extended time under anesthesia does impact the patient.Further attempts we made to get additional information but none was provided.This report is being raised on the basis of device malfunction with potential for injury upon reoccurrence.
|
| |
|
Event or Problem Description
|
|
Additional information received: the customer reported that the c8026, bioscrew hyperflex guidewire, broke during surgery.This caused a 2 hour delay, there was no reported patient injury, however, the extended time under anesthesia does impact the patient.Further information was received on (b)(6) 2018 due to a medwatch.This new information explained the 2 hour delay; part of the guidewire was broken off into the allograft.Flouroscopy was used to remove the broken piece, but by doing so, it also destroyed the allograft.The surgeon had to replace the allograft with a new one.The surgeon used x-ray to confirm all the pieces were retrieved.This report is being raised on the basis of device malfunction with potential for injury upon reoccurence.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections: was previously 10/2/18 corrected to 9/26/18 udi (b)(4).This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to b5, due to new information provided to conmed via a medwatch report number mw5079894.This did not change any of the evaluation reults or patient outcome.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
Additional information received: the customer reported that the 5129491, bioscrew hyperflex guidewire, broke during surgery.This caused a 2 hour delay, there was no reported patient injury, however, the extended time under anesthesia does impact the patient.Further information was received on (b)(6), 2018 due to a medwatch.This new information explained the 2 hour delay; part of the guidewire was broken off into the allograft.Flouroscopy was used to remove the broken piece, but by doing so, it also destroyed the allograft.The surgeon had to replace the allograft with a new one.The surgeon used x-ray to confirm all the pieces were retrieved.This report is being raised on the basis on device malfunction with potential for injury upon reoccurence.
|
| |
|