• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE-R READY-SCREEN (3); REAGENT RED BLOOD CELLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE-R READY-SCREEN (3); REAGENT RED BLOOD CELLS Back to Search Results
Catalog Number 0066814
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The immucor laboratory tested retention product on (b)(6) 2018 which performed as expected.The immucor laboratory also tested a returned blood sample from the customer site on (b)(6) 2018 (sample number (b)(6)) against retention product of (b)(4) which yielded an unexpectedly negative outcome, but with visually slight adherence around the button with both well 2 and 3.Therefore the customer issue with the blood sample testing was reproduced overall.No errors were reported during customer testing.The internal immucor record number is (b)(4).
 
Event or Problem Description
On (b)(6) 2017, a european french customer reported an unexpectedly negative antibody screen with capture-r ready-screen (3), when tested on a galileo echo instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-SCREEN (3)
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key7933905
Report Number1034569-2018-00222
Device Sequence Number930758
Product Code KSZ
UDI-Device Identifier10888234001461
UDI-Public10888234001461
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
102707/5045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date11/14/2018
Device Catalogue Number0066814
Device Lot NumberE311
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2018
Initial Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
-
-