| Model Number |
10605 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device is a combination product.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2018.It was reported that crossing difficulties were encountered.A 2.25 x 28 synergy drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with a different device.No patient complications were reported.However, returned device analysis revealed stent damage. synergy ii us mr 2.25 x 28 mm stent delivery system was returned for analysis.A visual examination of the stent found proximal stent damage, stent struts were pushed and bunched distally leaving a section of the proximal balloon body exposed.The undamaged distal section stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|