• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product replaced on ran (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.
 
Event or Problem Description
Consumer reported complaint for low blood glucose test results.Customer is concerned with tests results of 93 and 73 mg/dl.Customer's expected fasting blood glucose test result range is 150 - 160 mg/dl.Customer did state that the truemetrix air meter had been dropped and that since that happened it had been reading low.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a blood test was performed by the customer non-fasting and produced test result of 201 mg/dl using truemetrix air meter.The product is stored according to specification in the living room.The test strip lot manufacturer's expiration date is 06/21/2019 and customer could not recall the open vial date.The meter memory was reviewed for previous test result history (customer could not see date/time): (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7934338
Report Number1000113657-2018-00998
Device Sequence Number2255217
Product Code NBW
UDI-Device Identifier00021292007447
UDI-Public(01)00021292007447
Combination Product (Y/N)N
Initial Reporter StateMI
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/21/2019
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMV2722
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2018
Initial Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-