| Catalog Number |
502-03-46B |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Product not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total hip arthroplasty (tha), a mis-seal of the inner blister package of the primary tritanium hemi cluster hole cup 46mm was found.A spare was used instead.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding alleged "packaging issue" involving a tritanium shell was reported.The event was confirmed.Method & results: product evaluation and results: the device packaging was not returned.However, an image was provided showing a half opened tyvek lid.No additional information can be determined from the image.Medical records received and evaluation: clinician review was not performed as there is no indication that the event was related to patient factors.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there have been no other events for the referenced lot.Conclusions: the reported event was confirmed.However, the root cause could not be determined as the device was not returned.An nc investigation documented the following "it is possible that while the tyvek was being cut by the r50, a piece of tyvek did not fully detach from the roll of tyvek as it was being pulled.Because the tyvek was not fully detached, the tyvek was possibly not fully flush with phalanges.The fold could have occurred when the hot plate sealed the not fully flush tyvek.While it is possible that this may have been the cause of the fold on the inner blister, it could not be fully confirmed".If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total hip arthroplasty (tha), a mis-seal of the inner blister package of the primary tritanium hemi cluster hole cup 46mm was found.A spare was used instead.
|
| |
|