| Model Number |
B33098 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A used b33098 trifuse extension set was returned with one of the lines separated at the bond between the tubing and the trifurcated adapter.Microscopic examination of the bond site revealed that there was no solvent applied at the bond between the tubing and the trifurcated bond pocket.The probable cause of the bond separation is no solvent applied during manual product assembly to the tube bond interface at the trifurcated adapter.No dhr lot review was conducted because no lot number(s) was/were identified.
|
| |
|
Event or Problem Description
|
|
The event involved a customer report stating that one lumen of a smallbore trifuse extension set w/3 microclave became detached at the hub allowing for free flowing backflow from the central line.There was no serious injury reported.
|
| |
|