| Model Number |
802112 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation is in progress, but not yet concluded.
|
| |
|
Event or Problem Description
|
|
Upon receipt of the device, the status of the pressure module remained orange status instead of changing to green when the pump was online.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint could not be confirmed.During laboratory analysis, the product surveillance technician (pst) observed the pressure module to function normally throughout evaluation.Per log analysis, the reported issue occurred on (b)(6) 2018, but the log does not show the module was powered up on that date.It was used on (b)(6) 2018 and the next power up was on (b)(6) 2018.There are no indications of an issue in the log.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|