MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120I HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120I HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE Back to Search Results
Model Number ADVIA 2120I HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE
Medical Device Problem Code Device Handling Problem (3265)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer contacted a siemens customer care center and reported that an operator was exposed to advia ez wash solution while performing maintenance on the advia 2120i hematology system with single aspirate when the solution bottle slipped.The customer did not provide details on how the solution bottle slipped or how the operator was exposed to the advia ez wash solution when the operator was wearing personal protective equipment (ppe).The advia 120/advia 2120/advia 2120i hematology systems operator's guide advises operators' to "wear facial protection, gloves, and protective clothing" when performing maintenance on the system.The cause of the operator's exposure to the advia ez wash solution is unknown.The system is performing according to specifications.No further evaluation of this device is required.
 
Event or Problem Description
An operator of an advia 2120i hematology system with single aspirate was exposed to advia ez wash solution while performing maintenance on the system; the advia ez wash solution splashed onto the customer's face and eye when the solution bottle slipped.The operator rinsed her eye and sought ophthalmic consultation.The physician(s) prescribed a treatment of antibiotic eye drops, lubricant moisturizing eye drops, eye drops for dry eyes and vitamin a ointment for a month to the operator.There are no known reports of adverse health consequences due to the operator's exposure to advia ez wash solution.
 
Additional Manufacturer Narrative
Siemens filed the initial mdr 2432235-2018-00370 on (b)(6) 2018.Additional information (05-oct-2018): the customer clarified that the operator was exposed to advia ez wash solution when the bottle slipped from the operator's hand.The bottle fell and hit the table.Due to this, contents from the bottle splashed onto the operator's face and eye.The system is performing according to specifications.No further evaluation of this device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA 2120I HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE
Common Device Name
ADVIA 2120I HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave
tarrytown NY 10591
MDR Report Key7935851
Report Number2432235-2018-00370
Device Sequence Number10955092
Product Code GKZ
UDI-Device Identifier00630414019789
UDI-Public00630414019789
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K162977
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberADVIA 2120I HEMATOLOGY SYSTEM WITH SINGLE ASPIRATE
Device Catalogue Number11219530
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/15/2018
Supplement Date Received by Manufacturer10/05/2018
Initial Report FDA Received Date10/04/2018
Supplement Report FDA Received Date10/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age23 YR
-
-