ZIMMER BIOMET, INC. G7 OSSEOTI 3 HOLE SHELL 50MM D; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Complaint sample was evaluated and the reported event was confirmed.Upon visual inspection the liner has been fractured.However, no further analysis was performed due to the impaction to the liner.There is circular gouging on the liner.There is no damage to the outside of the shell however the inside diameter has deep gouging and damages to the locking feature.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2018-04137.
 
Event Description
It was reported that during surgery, the liner would not seat into the shell.The shell was removed and the surgery was finished with a second shell and liner.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI 3 HOLE SHELL 50MM D
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7936351
MDR Text Key122793869
Report Number0001825034-2018-09442
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK140669
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110010243
Device Lot Number6234846
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-