Model Number 000714
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter immediately came off after placement.It was confirmed that sterilized water leakage was found from the balloon when the balloon was inflated again and checked.There was no reported patient injury.
 
Event or Problem Description
It was reported that the catheter immediately came off after placement.It was confirmed that sterilized water leakage was found from the balloon when the balloon was inflated again and checked.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records (dhr) review is not available because the corporate lot number is unknown.Investigation summary: the sample returned for evaluation was one 14 fr tri-funnel replacement gastrostomy feeding tube.Visual, microscopic, and functional evaluations were performed of the physical sample.The sample is returned with a small pinhole leak emanating from the mid section of the balloon.Examination of the leak site revealed two superficial external surface damage slits, emanating from the outside inward which did not penetrate through the entirety of the wall, except for a resulting small hole in the balloon wall of one of the slits.The damage is indicative of sharp instrument contact but it could not be determined when/where that contact occurred, whether it was preexisting or occurred during/following insertion.This investigation is unconfirmed for deflation, cause unknown and confirmed for pinhole leak.The investigation indicates that the failure is resulting from pinhole leak, but the exact cause is unknown.Labeling review: a review of the product instructions for use (ifu) procedural instructions, indications, warnings, precautions, possible complications and contraindications showed that the product labeling is adequate.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 14F
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7936391
Report Number3006260740-2018-02713
Device Sequence Number1514571
Product Code KNT
UDI-Device Identifier00801741037122
UDI-Public(01)00801741037122
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000714
Device Catalogue Number000714
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer01/27/2019
Initial Report FDA Received Date10/04/2018
Supplement Report FDA Received Date02/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-