• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS FULL KIT 18G X 10 CM WITH BIOPATCH AND PROBE COVER; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS FULL KIT 18G X 10 CM WITH BIOPATCH AND PROBE COVER; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number F318108PT
Device Problems Positioning Failure (1158); Material Puncture/Hole (1504); Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/01/2018
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of guidewire damage and difficult device removal was confirmed and the cause appeared to be use-related.The product returned for evaluation was one 18ga x 10cm powerglide pro rt midline catheter assembly.Blood residue was observed on the sample.The catheter was advanced on the needle shaft and the safety mechanism was engaged over the needle bevel.The housing was disassembled.The distal end of the guidewire was detached and was received loose.The weld tip appeared intact.The catheter exhibited curved shape memory and the tip appeared to exhibit deformation.Microscopic inspection of the broken wire fragment confirmed that the weld tip was intact.Inspection of the break in the core wire revealed a granular fracture surface.The fracture exhibited a region of increased luster.Curved shape memory was observed in the vicinity of the break.Inspection of the needle bevel revealed abundant ridged and outward flaring deformation along the proximal edge.Inspection of the distal tip of the catheter confirmed deformation and discoloration.The edges of the tip curved inward.The region of increased luster in the core wire and the needle bevel damage were consistent with wire damage initiated by withdrawal against the needle bevel.The catheter damage was also consistent with guidewire withdrawal.The blood residue indicated that damage occurred during attempted device insertion.A lot history review (lhr) of resc1192 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported by the facility that the rn deployed wire with no resistance but could not fully deploy catheter.It was stated there was some movement forward of the ¿wings¿, but would not move it forward into the vein.The rn tried to pull back wire but was not able to do so.The entire device was removed from patient, but again it was stated the rn encountered resistance.It was reported that when removed it was noted the wire had punctured catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL KIT 18G X 10 CM WITH BIOPATCH AND PROBE COVER
Type of Device
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key7936753
MDR Text Key123067119
Report Number3006260740-2018-02727
Device Sequence Number0
Product Code FOZ
Reporter Country CodeUS
PMA/PMN Number
K162377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberF318108PT
Device Catalogue NumberF318108PT
Device Lot NumberRESC1192
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2018
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received 09/10/2018
Initial Date FDA Received10/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-