MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3542715 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Concomitant products: mentor siltex contour profile high 650cc saline catalog: 3542715 lot: 5931916 sn: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) -year old caucasian female patient underwent primary breast augmentation with mentor siltex contour profile high 650cc saline breast prostheses.Deflation of the right breast prosthesis was determined.As a result, the patient underwent removal on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
On 11/21/2018, the device investigation summary was updated to: upon receipt by mentor, the device contained no fluid.During initial evaluation siltex cracking was observed on the posterior aspect.Lines of shell wear were also observed on the posterior aspect, suggesting in-vivo folding or creasing of the device.Leak testing of the device, in accordance with mentor procedures, revealed four rents within the siltex cracking, measuring approximately rent a 0.6 cm, rent b 0.3 cm, rent c 0.2 cm and rent d 0.4 cm.No other anomalies were observed.Complaint was confirmed.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.Mentor product analysis lab concluded that the rent, siltex cracking and shell wear occurred sometime subsequent to the removal of the device from its protective packaging.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Mentor believes that siltex cracking is ultimately a result of high stresses caused by acute folds (i.E., severe ¿v¿ or tight folds) in the room temperature vulcanization (rtv) shell of siltex breast implants.The device history record (dhr) of lot number 5931916 was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 10/17/2018, the mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device contained no fluid, upon receipt by mentor.No foreign material was observed within the device or on the shell surface.During visual examination, mentor product analysis lab observed the device appears intact.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, mentor product analysis lab was precluded from further evaluating the device.No anomalies were found.The complaint of deflation was not confirmed since no anomalies were found on the reported device.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|