OLYMPUS MEDICAL SYSTEMS CORP. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VHR
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  Malfunction  
Manufacturer Narrative

The device was returned to olympus for evaluation. A visual inspection was performed on the returned device and the bending section cover glue from the distal end and insertion tube side cracked with portions of the cement missing/broken off. The missing/broken off cement was not returned. There were scratches noted on the bending section cover glue, which likely cause the damage to the cement. The scope¿s biopsy channel was inspected with olympus telescope and no missing parts or foreign objects/materials were noted inside the channel. Based on the reported complaint the cause of the damage on the bending section cover cement is due to mishandling the instruction manual provides warning which states, ¿do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector. Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force. The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations. It could also cause parts of the endoscope to fall off inside the patient. ¿.

 
Event Description

Olympus was informed that during an unspecified procedure, a piece of the scope¿s distal tip fell inside the patient's bladder. It is unknown if the device fragment was retrieved or if the intended procedure was completed.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCYSTO-NEPHRO VIDEOSCOPE
Type of DeviceCYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
4089355124
MDR Report Key7937063
MDR Text Key123201997
Report Number2951238-2018-00607
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/01/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/04/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-VHR
Device Catalogue NumberCYF-VHR
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer09/27/2018
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/13/2019
Was Device Evaluated By Manufacturer? No
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-